Title: | Systematic Studies of a Sapphire Bolometer with Phonon Pulses in the Temperature Range of 10–100mK |
Authors: | Garai, A. Mazumdar, A. Reza, A. Krishnmoorthy, H. Gupta, G. Nanal, V. Pillay, R.G. Ramakrishnan, S. |
Keywords: | NTD Ge sensor Cryogenic bolometer Vibrational noise |
Issue Date: | 22-Dec-2020 |
Abstract: | Systematic studies with a sapphire bolometer test setup, made with indigenously made NTD Ge sensor, are carried out. A C++ and ROOT-based pulse analysis program implementing Savitzky–Golay filtering technique for analyzing the bolometer signal is discussed. Response of the bolometer subject to heater pulses of energy equivalent of 0.3–5MeV is presented in the temperature range of 10–100mK. The energy resolution is found to be 15±3 keV in the temperature range of 15–25mK. It is observed that the energy resolution of the bolometer is nearly constant ∼ 15±3 keV over the energy range of 0.3–5MeV. Performance of the bolometer with the addition of a moderatesize tin sample (∼0.6 g) to the sapphire substrate is presented. The impact of vibration on the bolometer is also discussed. |
URI: | http://localhost:8080/xmlui/handle/123456789/1702 |
Appears in Collections: | Year-2020 |